Getinge

Micro-matrix Bioreactors (1-10 ml)

The Micro-Matrix allows control of up to 24 independent bioreactors in a single microtiter plate area.
For each of the 24 bioreactors with a volume of 1-10 ml on the plate, there is independent control of the following parameters, just like larger scale bioreactors:

• pH (measurement and double-sided control)
• temperature (measurement and two-sided control, possibility to create gradient across the whole plate)
• dissolved oxygen control (measurement and bilateral control)
• individual liquid addition (including feed profiles)
• up to 4 different gas additions, each individually controlled

Key Features

  • Minimum space requirement in laboratory
  • Monitoring and control of pH and dissolved oxygen
  • Separate control of wells
  • Parallel operation of 24 bioreactors

Application Areas

  • Screening of cell line, microbial or yeast libraries,
  • Process development studies,
  • Process optimization studies,
  • Small volume culture studies.

The Micro-Matrix allows control of up to 24 independent bioreactors in a single microtiter plate area.

For each of the 24 bioreactors with a volume of 1-10 ml on the plate, there is independent control of the following parameters, just like larger scale bioreactors:

  • pH (measurement and two-sided control)
  • temperature (measurement and two-sided control, possibility to create gradient across the whole plate)
  • dissolved oxygen control (measurement and two-sided control)
  • individual liquid addition (including feed profiles)
  • Up to 4 different gas additions, each individually controlled

Software

Tape

Each well on the cassette has a separate pH and oxygen sensor and each well has 3 temperature sensors mounted under the cassette, guaranteeing optimum temperature measurement and control.

General:

  • Number of wells per cassette 24
  • Total volume per bioreactor well 10 mL
  • Working volume per bioreactor well 1 - 5 mL

Mixing:

  • Orbit radius 25 mm diameter
  • RPM range 0 - 400 RPM
  • Mixing times < 1 sec

Aeration:

  • Gas delivery 0.1 - 18 mL/min Per gas, per well
  • Number of gasses 4
  • Gas feed pressure > 1.0 (extern)

Liquid additions:

  • Liquid delivery > 50 nl
  • Number of liquids is 1 per 6 wells
  • Liquid feed pressure > 2.0 bar (extern)

Temperature:

  • Measurement range 0 - 45 °C
  • Resolution 0.1 °C
  • Accuracy 0.5 °C
  • Temperature control range 15 - 45 °C
  • ∆T between adjacent wells 1 °C
  • Heat-up time 1 °C / min 20 °C to 37 °C
  • Control accuracy ± 0.1 °C

pH:

  • Measurement 5.5 - 8.5 pH
  • Resolution < 0.05 pH @ pH 7.2
  • Accuracy < 0.1 pH @ pH 7.2
  • pH control range 5.5 - 8.5 pH
  • Control accuracy ± 0.1 pH

Dissolved Oxygen:

  • Measurement 0 - 150 % Air saturation
  • Resolution < 1 % Air saturation
  • Accuracy < 0.5% @ 0% air saturation
  • < 3% @ 100% air saturation
  • DO control range 0 - 150 %
  • Control accuracy ± 5% Air saturation
  • Oxygen Transfer Rate 250 mmol / L / h

Dimensions:

  • 415 mm Height
  • 570 mm Depth
  • 560 mm Width

Weight:

65 kg