Labconco XPert Weigh Boxes are glove boxes designed for use in pharmaceutical research, nanotechnology, biochemistry and in cleanroom applications requiring ISO Class 3 conditions. Their inlet and outlet HEPA filters provide a leak-tight physical barrier to protect the operator from hazardous airborne particulates and powders. Their uses include handling hazardous particulates and powders, microorganisms, low-level radiochemicals, carcinogens and asbestos. These isolators include an anti-static ionizer fan to reduce static charge and a balance vibration isolator, both useful during weighing operations. These polyethylene-lined ventilated glove boxes feature HEPA filters, anti-static ionizer fan and a balance vibration isolator for weighing operations.
One-piece molded, medium-density polyethylene shell. Labconco recommends this website for a comprehensive listing of chemical resistance data for polyethylene.
Dry powder epoxy-coated steel superstructure
1/4" thick laminated safety glass viewing window, 26.4" wide x 12.5" high, removable for loading equipment, with powder-coated steel frame and neoprene gasket
Space-saving inner and outer transfer chamber doors that pivot upward, are counterbalanced and equipped with quick-latches
Black, chemical-resistant work surface sealed to the glove box floor
Anti-static ionizer fan to circulate ionized air inside the box to reduce static charge during weighing
Interior ceiling-mounted electrical duplex receptacle with 5 amp maximum circuit load
Neoprene gloves, size 9 3/4", are included.
Built-in blower with speed control capable of airflow from 13 CFM to 75 CFM.
Fluorescent lamp, 22 watts, with easy replacement from the top
Switches for light, blower and innermost interior electrical receptacle
Inlet and outlet HEPA filters, 99.99% efficient on 0.3 micron particulates
Test port for challenging HEPA filter integrity
Main chamber static pressure gauge for monitoring HEPA filter loading.
Approximately 13 cubic feet of interior space in main chamber
11.0" nominal ID x 12.0" long transfer chamber
Factory leak tested with a mass spectrometer while glove box is pressurized with helium at 5 inches of water gauge
No detectable leaks greater than 1 x 10(-3) ml/sec
Meet Class 1 atmosphere containment conditions for oxygen leak rates per ISO 10648-2 test method
Transfer chamber interior interlocking shelf of dry powder epoxy-coated steel
0.05 ppm average tracer gas levels or less detected during perimeter scan testing (surpasses ASHRAE 110 standards)
Airborne particulate cleanliness (measured under positive and negative pressures per ISO 14644-1 test method) exceeds ISO Class 5 conditions (equivalent to Class 100) and achieved ISO Class 3 conditions (equivalent to Class 1) at both minimum and maximum operational airflow volumes
Vibration less than 1.5 x 10(-5) inches displacement measured at the work surface at both minimum and maximum airflow volumes